Model Number 2088TC/58
Device Problems Failure to Capture (1081); Device Sensing Problem (2917); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2024
Event Type  Injury  
Event Description
It was reported that the right ventricular (rv) lead dislodged.Upon further interrogation, it was noted that the rv lead exhibited failure to capture and reduced r-wave sensing.The rv lead was explanted and replaced.The patient was in stable condition.
 
Manufacturer Narrative
The reported events were lead dislodgement, loss of capture and r-wave amp variation.A complete lead was returned in one piece with the helix retracted and clogged blood/tissue.After cleaning and applying the torque directly to the connector pin, the helix could be extended and retracted.The full helix extension length was measured to be within specification.The reported events of loss of capture and r-wave amp variation were not confirmed.Electrical testing was normal with no anomalies found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18956756
MDR Text Key338305129
Report Number2017865-2024-35616
Device Sequence Number1
Product Code NVN
UDI-Device Identifier05414734502894
UDI-Public05414734502894
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2088TC/58
Device Lot NumberA000136474
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/04/2024
Initial Date FDA Received03/21/2024
Supplement Dates Manufacturer Received03/22/2024
Supplement Dates FDA Received03/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ASSURITY MRI PACEMAKER.; TENDRIL LEAD.
Patient Outcome(s) Required Intervention;
Patient Age83 YR
Patient SexFemale
-
-