Model Number 96570-115
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/19/2024
Event Type  malfunction  
Event Description
Medtronic received information that during use of a bio-medicus nextgen arterial cannula , the customer reported that while the device was being inserted into the internal jugular vein on a patient undergoing a minimally invasive procedure the customer reported that the tip completely caved in, which is abnormal for a cannula.The customer attempted to re-open it tried to straighten it out with an introducer but the tip still had a visible kink.The defect was observed before the cannula was inserted into the internal jugular.The device was replaced to complete the procedure.There was no adverse patient effect associated with this event.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Manufacturer Narrative
Medtronic received additional information that the cannula was not heated or cooled prior to use.The customer used the initial insertion site when the replacement cannula was inserted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Manufacturer Narrative
Device evaluation summary: visual inspection shows evidence of damage to the tip.Reason for return was confirmed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO-MEDICUS NG ARTERIAL CANNULA
Type of Device
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18957571
MDR Text Key338469623
Report Number9612164-2024-01395
Device Sequence Number1
Product Code DWF
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K142673
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 05/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number96570-115
Device Catalogue Number96570-115
Device Lot Number226450696
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/20/2024
Initial Date FDA Received03/22/2024
Supplement Dates Manufacturer Received04/10/2024
05/02/2024
Supplement Dates FDA Received05/02/2024
05/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-