• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number X SERIES
Device Problem Sparking (2595)
Patient Problem Necrosis (1971)
Event Date 02/29/2024
Event Type  Death  
Event Description
Complainant alleged that while attempting to defibrillate a 70-year-old male patient, sparking was seen from the associated defib pad.Complainant indicated that the patient subsequently expired, however it was not a result of the reported malfunction.
 
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Manufacturer Narrative
The device and multifunction cable (mfc) were returned to zoll medical corporation for evaluation.The defib pads used during the event were not returned for evaluation.The device was put through extensive testing using the customer's returned mfc without duplicating the report.Review of the device log shows a large difference between the measured impedances which is evidence of poor coupling between the patient and the electrodes being used.The first discharge had a discharge impedance of 106 ohms and delivered 256.3 joules.The second discharge had a discharge impedance of 217 ohms and delivered 304 joules.This second shock resulted in the customer's report of the pad sparking.Despite the shock, the discharge appeared to have been successful which was further confirmed through communication with the customer.It was confirmed that therapy was delivered to the patient.Electrode labeling states the importance of good placement on the patient and provides instruction for proper electrode application technique.Zoll recommends that patients are cleaned and hair is clipped prior to applying electrode pads to assure good coupling of electrode to skin contact.Poor adherance and/or air under the electrodes can lead to the possibility of arcing/sparking.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key18958288
MDR Text Key338341525
Report Number1220908-2024-01090
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00847946019259
UDI-Public00847946019259
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/04/2024
Initial Date FDA Received03/22/2024
Supplement Dates Manufacturer Received03/04/2024
Supplement Dates FDA Received04/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age70 YR
Patient SexMale
-
-