Model Number 7122Q/65
Device Problems Difficult to Fold, Unfold or Collapse (1254); Use of Device Problem (1670); Device Dislodged or Dislocated (2923); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the patient's right ventricular lead was replaced during implant due than an unknown malfunction.An alternative lead was implanted on (b)(6) 2024.The patient's condition was unknown.
 
Event Description
Additional information received noted that the right ventricular lead dislodged and could not be implanted during procedure.It was also noted that the helix of the lead felt as if some tissue was stuck in it.
 
Manufacturer Narrative
The reported event of helix mechanism issue was confirmed while the reported event of unable to implant and lead dislodgement were not confirmed.As received, a complete lead was returned in one piece.For analysis.Visual inspection found the helix extended, bent, stretched, and clogged with blood.X-ray examination found the inner coil over-torqued consistent with procedural damage.The cause of helix mechanism issue was isolated to the helix being damaged, clogged with blood, and over-torque of the inner coil.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18959638
MDR Text Key338718386
Report Number2017865-2024-35619
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734503341
UDI-Public05414734503341
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7122Q/65
Device Lot NumberA000142727
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/29/2024
Initial Date FDA Received03/22/2024
Supplement Dates Manufacturer Received03/28/2024
Supplement Dates FDA Received04/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age78 YR
Patient SexMale
-
-