A customer contacted the binding site (pdx) on the 08th of february 2024 to report that their pdx freelite® human kappa free kit for use on the roche cobas® c systems assay (product code: lk016.Cb, kit lot: 490912) was producing unusually high kappa results on one of their instruments.On the (b)(6) 2024, the customer informed pdx that 20 results released from the laboratory for kappa free were falsely elevated.15 of these results were above the total allowable error (tea) of 40% of the assay.The testing took place on the (b)(6) 2024, (b)(6) 2024 and (b)(6) 2024 to the (b)(6) 2024 and the discrepancy was identified by the clinician on (b)(6) 2024.Upon repeat testing on an additional analyser, the customer obtained results which they reported to be more in line with the clinical picture.The customer reported that for the initial instrument the issue was resolved following replacement of a sample probe and a new kit lot.Although no reports have been received of harm to patients, change in patient management or inappropriate therapy, the decision was taken to report as a matter of caution.If the error were to recur for 1 of the 15 released discrepant results above the tea there is a remote possibility of it leading to serious injury.For this sample the falsely elevated free light chain (flc) kappa value was 643 mg/l whereas on repeat the value returned was 0.63mg/l (below the reference range; 3.30 - 19.40 mg/l).Should the error recur there is a potential for patient management to be changed, if the assay result is used in isolation.In a diagnostic scenario, this issue could lead to a false diagnosis of myeloma and subsequent treatment on the basis of the flc criterion of the slim criteria (involved/uninvolved flc equal to or greater than 100, provided that the involved flc is equal to or greater than 100mg/l) in the presence of more than 10% bone marrow plasma cells.As per the instructions for use of the assay (sin129), results of this assay should always be assessed in conjunction with the patient's medical history, clinical examinations, and other findings including previous freelite results if available.Case investigation is continuing.
|