• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES BD FACS¿ SAMPLE PREP ASSISTANT III; SEE H.10

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES BD FACS¿ SAMPLE PREP ASSISTANT III; SEE H.10 Back to Search Results
Catalog Number 647205
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/28/2024
Event Type  malfunction  
Event Description
It was reported that while using bd facs¿ sample prep assistant iii erroneous results were obtained.No patient impact.The following information was provided by the initial reporter: customer reported the spa is not adding reagent to the sample tube.A full rack of samples was prepped, the first 12 looked and ran fine, but the remaining 28 did not have any reagents in them (erroneous results).They discarded the affected samples and prepped the primary samples again on another spa where they processed successfully.Samples were improperly prepared by the spa and when they were run on the cytometer they produced erroneous patient results, which were noticed by the lab technicians who discarded the samples and re-prepped new ones on another spa.
 
Manufacturer Narrative
D2a: common device name: automated pipetting, diluting and specimen processing workstations for flow cytometric analysis.H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD FACS¿ SAMPLE PREP ASSISTANT III
Type of Device
SEE H.10
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
2350 qume drive
san jose CA 95131
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
2350 qume drive
san jose CA 95131
Manufacturer Contact
fahmy razak - mdr
2350 qume drive
san jose, CA 95131
4089542435
MDR Report Key18961262
MDR Text Key338991310
Report Number2916837-2024-00024
Device Sequence Number1
Product Code PER
UDI-Device Identifier00382906472052
UDI-Public(01)00382906472052
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131301
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number647205
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/28/2024
Initial Date FDA Received03/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/22/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-