MPRI SPRINT QUATTRO SECURE MRI SURESCAN; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number 6947M62
Device Problems Signal Artifact/Noise (1036); Fracture (1260); High impedance (1291); Over-Sensing (1438)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/14/2024
Event Type  Injury  
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported that the right ventricular (rv) lead exhibited undefined high pacing impedance, noise, and triggered a lead integrity alert (lia) due to high rate non-sustained episodes and sensing integrity counter (sic).Fracture was confirmed.The lead was programmed off and was later capped and replaced. no patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated oversensing associated with the right ventricular lead.Analysis of the device memory indicated the criteria for the right ventricular lead integrity alert were met.Analysis of the device memory indicated the impedance of the right ventricular pacing lead was beyond the expected upper range.Analysis of the device memory indicated oversensing due to non-physiologic signals/sensing integrity counter.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE MRI SURESCAN
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18964691
MDR Text Key338430391
Report Number2649622-2024-07994
Device Sequence Number1
Product Code NVY
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P920015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/01/2013
Device Model Number6947M62
Device Catalogue Number6947M62
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/14/2024
Initial Date FDA Received03/22/2024
Supplement Dates Manufacturer Received03/27/2024
Supplement Dates FDA Received04/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/04/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DVFB2D4 ICD
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age69 YR
Patient SexMale
-
-