MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CONTINU-FLO CLEARLINK SYSTEM SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CONTINU-FLO CLEARLINK SYSTEM SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2R8537
Device Problems Fluid/Blood Leak (1250); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2024
Event Type  malfunction  
Event Description
It was reported that the tubing separated from the y-stie connection of a continu-flo clearlink system solution set which resulted in a leak.This incident occurred after priming and prior to dispensing.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Additional information was added to h3, h4, h6, and h10: h10: the actual device was not available; however, a photograph of the sample was provided for evaluation.The returned photograph was reviewed, and it was noted that there was a separation between the tubing and the y-site clearlink.The reported condition was verified.The cause of the reported condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTINU-FLO CLEARLINK SYSTEM SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18965051
MDR Text Key338447573
Report Number1416980-2024-01296
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412565743
UDI-Public(01)00085412565743
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K180739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2R8537
Device Lot NumberR23H24017
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/26/2024
Initial Date FDA Received03/22/2024
Supplement Dates Manufacturer Received04/24/2024
Supplement Dates FDA Received04/25/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-