This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause of the observed insertion tube coating peeling issue could not be determined, however, the issue was likely due repetitive use stress, external factors and or user handling/mishandling.The event may be detected/prevented by following the instructions for use (ifu) section below: lf-tp/dp/gp instruction manual "chapter 3 preparation and inspection 3.2 endoscope preparation and inspection.Olympus will continue to monitor field performance for this device.
|