SMITH & NEPHEW, INC. REF BALL JNT SCREWDRIVER SHAFT Back to Search Results
Catalog Number 71362295
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/29/2024
Event Type  malfunction  
Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Event Description
It was reported that, during a thr surgery, the tip of a ref ball jnt screwdriver shaft broke off.The procedure was resumed, without any delay, using a s+n back-up device.No injury was reported as a consequence of this issue.
 
Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device reveals the screwdriver, spring seat and spring broke off.These pieces were not returned.The device shows signs of extensive wear and use.This inspection was conducted by naked eye.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REF BALL JNT SCREWDRIVER SHAFT
Type of Device
SCREWDRIVER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18965218
MDR Text Key338467508
Report Number1020279-2024-00588
Device Sequence Number1
Product Code HXX
UDI-Device Identifier03596010454201
UDI-Public03596010454201
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71362295
Device Lot Number10CM12539
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2024
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/29/2024
Initial Date FDA Received03/23/2024
Supplement Dates Manufacturer Received04/19/2024
Supplement Dates FDA Received04/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/19/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-