• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ESSENTIO DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ESSENTIO DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number L101
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Telemetry Discrepancy (1629); Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2024
Event Type  Injury  
Event Description
It was reported that the patient implanted with this pacemaker had observed yellow collecting waves on their communicator, and patient-initiated interrogation (pii) were unsuccessful.A few days later, it was determined that this pacemaker had entered safety mode.Upon interrogation, the following codes were noted: 0x12 0xf 0xe.Technical services (ts) discussed troubleshooting options, recommending emergent device replacement and product return for analysis.Additional information received reported that this pacemaker was explanted and replaced with another manufacturer's device.No additional adverse patient effects were reported.This product is expected to be returned by the explanting facility.
 
Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESSENTIO DR
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18966071
MDR Text Key338430794
Report Number2124215-2024-17635
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526558924
UDI-Public00802526558924
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
N970003/S167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial
Report Date 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/26/2020
Device Model NumberL101
Device Catalogue NumberL101
Device Lot Number768082
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/04/2024
Initial Date FDA Received03/23/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/22/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age76 YR
Patient SexMale
-
-