C.R. BARD, INC. (BASD) -3006260740 POWER PORT MRI ISP, 8 FR GROSHONG, INT WO SP, SL; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 8808561 |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/26/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that during the preparation of a port placement procedure, the lock connection was allegedly loose.The procedure was completed using another device.There was no patient contact.
|
|
Manufacturer Narrative
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a video was provided for review.The investigation of the reported event is currently underway.H10: d4 (expiration date: 05/2025) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|
Event Description
|
It was reported that during the preparation of a port placement procedure, the lock connection was allegedly loose.The procedure was completed using another device.There was no patient contact.
|
|
Manufacturer Narrative
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned.One electronic video was provided for review.The cathlock is noted to be loose.Therefore, the investigation is confirmed for the reported loose connection.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 05/2025), g3, h6 (method) h11: h6 (result, conclusion) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|