Model Number 2ACH20
Device Problems Material Deformation (2976); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2024
Event Type  malfunction  
Event Description
It was reported that during a cryo ablation procedure, after inflation was performed and the right superior pulmonary vein (rspv) was frozen.The mapping catheter shaft inside the balloon catheter was slightly curved under imaging. the case was completed with cryo.After the case, the mapping catheter was removed and it was confirmed that the mapping catheter tip was kinked. no patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Manufacturer Narrative
Product event summary: the 2ach20 mapping catheter with lot number 8812882 was returned and analyzed.Visual inspection of the shaft segment area showed the shaft was kinked approximately 57.6 inches from the lemo connector.The functional test was performed and the continuity and impedance measurement between the electrodes and the other side of the cable showed the electrode's continuity and impedance to the cable are normal.In conclusion, the reported kink issue was confirmed and the mapping catheter failed the returned product inspection due to a shaft kink.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE ADVANCE MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18967109
MDR Text Key338439209
Report Number2182208-2024-01256
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K162892
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2ACH20
Device Catalogue Number2ACH20
Device Lot Number8812882
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/11/2024
Initial Date FDA Received03/25/2024
Supplement Dates Manufacturer Received04/02/2024
Supplement Dates FDA Received04/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/26/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-