The philips engineer reviewed the ecg trace, revealing the patient's pacemaker continued to generate stimulation peaks and simulated beats at the time the monitor measured a heart rate of 70 bpm; therefore, no ecg alarm was generated.Desaturation and pulse alarms were occurring at this time as well.Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.(b)(6).
|
It was reported the intensive care unit (icu) patient was found deceased but no ecg alarms for bradycardia or asystole were generated at the monitor or central station.When the biomed reviewed the logs, it was determined there were no ecg alarms during the event, with the heart rate being around 70bpm wit pvcs.It was also indicated only desaturation alarms of <65% were noted.
|
A clinical product specialist reviewed the clinical audit log and the rhythm strip provided.On february 24, 2024, at 08:44:04, the spo2 low limit was changed to 83 at the bedside monitor.At 08:44:08, the desat limit was changed to 80 at the bedside monitor.At 08:44:10, the spo2 low limit was changed to 80 at the bedside monitor.Between 09:15 to 09:33, there are several desat alarms logged.The alarms were acknowledged at the bedside monitor.At 09:33:38, the desat limit was changed to 70, and the spo2 low limit was changed to 70.At 09:34:12, the respiratory and nbp alarms were switched off.When individual alarms are switched off, the ¿alarms off¿ symbol is shown beside the measurement numeric at the bedside and at the pic ix.No red or yellow physiological alarms are sounded, and no messages are shown.However, inop messages are shown, but the inop tones are not sounded.At 9:45:30, the desat limit was changed to 65.At 12:15:16, a desat alarm generated and ended at 12:42:30.Starting at 12:39, there are multiple ¿spo2 no pulse¿ inops.The alarms were acknowledged at the bedside at 12:43:51 and 12:49:58.At 12:50:31, the spo2 alarms were switched off.The monitor was put into standby at 12:50:55.A review of the rhythm provided at 12:50:46 shows a ventricular paced rhythm.There is a question mark for the spo2 and pulse numeric due to the ¿spo2 no pulse¿ inop logged in the audit log.The heart has two components: electrical and mechanical.The star algorithm processes the patient¿s electrical ecg signal to identify and classify the patient¿s beats, and then generate alarms.The electrical signal is displayed by the ecg waveform and the hr value.The star algorithm does not process the mechanical contraction of the heart which is displayed by the pulse value (from the spo2 or invasive pressure measurement) on the monitor.There is a clinical condition called pulseless electrical activity (pea).When the patient is in pea, the heart¿s electrical system continues to work but may be too week to stimulate the heart to contract, leading to cardiac arrest.The star algorithm is a tool to assist the clinician in managing and evaluating their patients.The rhythm strips provided continue to show the heart¿s electrical system responding to the pacer pulse resulting in no ecg alarms.Philips does not have an alarm for pea, because the star algorithm continues to recognize and classify the rhythm and display the hr value.The first alarm that you would get on the monitor in this condition is the ¿spo2 no pulse¿ inop.In conclusion, the patient's pacemaker continued to stimulate the heart at 70 bpm and the spo2 measurement displayed a question mark and a no pulse inop.This indicated the patient was in pea.The heart continued to respond to the paced pacemaker impulse, resulting in no ecg alarms.Philips does not have an alarm for pea and the device functioned as specified.Based on this information, the device did not cause or contribute to the reported event and functioned as designed.In addition, caregivers were changing alarm limits for desaturation over time, which suggests they were aware of the patient's deteriorating condition.Based on the information available, the device performed according to specification, and the criteria for ecg alarm generation was not met.The device remains in use at the customer site.
|