BOSTON SCIENTIFIC NEUROMODULATION VERCISE GENUS; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
|
Back to Search Results |
|
Model Number DB-1232 |
Device Problem
Inappropriate or Unexpected Reset (2959)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/19/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that the database analysis performed identified a bluetooth fault code when charging the deep brain stimulation (dbs).Noise from the charging field of the charger interferes with the bluetooth communication chipset, thus causing internal resets where the device stimulation momentarily turns off and then returns to normal operation.No other anomalies were found with the analysis as the device displayed a normal charge log.The ipg remains implanted in the patient and delivering therapy.
|
|
Event Description
|
It was reported that the database analysis performed identified a bluetooth fault code when charging the deep brain stimulation (dbs).Noise from the charging field of the charger interferes with the bluetooth communication chipset, thus causing internal resets where the device stimulation momentarily turns off and then returns to normal operation.No other anomalies were found with the analysis as the device displayed a normal charge log.The ipg remains implanted in the patient and delivering therapy.
|
|
Search Alerts/Recalls
|
|
|