Model Number LTF-190-10-3D
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
A review of the device history record found no deviations that could have caused or contributed to the reported issue.A definitive root cause of the foggy image was likely due to damage in the charged coupled device, however a definitive root cause could not be identified.The customer may be able to reduce and prevent occurrence of the event by handling device in accordance with the following ifu.If additional information becomes available at a later date, this report will be supplemented.Olympus will continue to monitor the field performance of this device.
 
Event Description
It was observed that during the device inspection, the deflectable videoscope exhibited foggy image due to bad charged coupled device (ccd) unit.There were no reports of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX 3D DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18970651
MDR Text Key338606641
Report Number9610595-2024-06315
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170434938
UDI-Public04953170434938
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
K123365
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberLTF-190-10-3D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/20/2021
Initial Date FDA Received03/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/08/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-