• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX SETS (OXIRIS); DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX SETS (OXIRIS); DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 112016
Device Problems Fluid/Blood Leak (1250); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/02/2024
Event Type  malfunction  
Manufacturer Narrative
Oxiris has been temporarily approved for use in the us under emergency use authorization eua (b)(4).With a specific indication to treat patients with covid-19 infection.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It is reported that an external fluid leak was observed from the filter and return line connection point of two (2) oxiris sets during setup/preparation.The line was "disconnecting from the filter easily".There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information: d9, h3, h6, h10.H10: two (2) devices were received for evaluation.Visual inspection of both sets showed that the screwed connector of the return line is broken inside the header blood port.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the observed damage was determined to be related to mechanical shock applied to the product during shipping, transport, and/or improper storage.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX SETS (OXIRIS)
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18970698
MDR Text Key338509139
Report Number8010182-2024-00113
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414012236
UDI-Public(01)07332414012236
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number112016
Device Lot Number22I0093CA
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/05/2024
Initial Date FDA Received03/25/2024
Supplement Dates Manufacturer Received04/17/2024
Supplement Dates FDA Received04/23/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-