• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAX MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAX MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955558
Device Problem Unexpected Shutdown (4019)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 02/20/2024
Event Type  Injury  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during continuous renal replacement therapy using a prismax machine, the machine "spontaneously and abruptly stopped".The treatment was terminated without returning the extracorporeal blood to the patient.It was further reported the screen was unresponsive.No patient symptoms were reported; however, the patient required a blood transfusion (two units).No additional information is available.
 
Manufacturer Narrative
H10: the actual device was not available.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.A review of the event history log confirmed the occurrence of the machine alarms which were compatible with the return line becoming clamped or kinked during use.The alarms were not successfully cleared by following operator manual procedure, which violated machine safety measures.The reported condition was verified.The cause of the condition was determined to be related to user error.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAX MACHINES
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - BROOKLYN PARK
7601 northland drive
n suite 170
minneapolis MN 55428
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18971150
MDR Text Key338512230
Report Number3003504604-2024-00005
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414124359
UDI-Public(01)07332414124359
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955558
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/28/2024
Initial Date FDA Received03/25/2024
Supplement Dates Manufacturer Received04/23/2024
Supplement Dates FDA Received04/26/2024
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-