Model Number N/A |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device will not be returned for analysis as it remains implanted; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted h3 other text : device location is unknown.
|
|
Event Description
|
It was reported that the surgeon had broached for a size 5 taperloc however when he implanted it, it sat very proud and would not go down.The surgeon removed the stem and re-broached for a size 5 with the broach countersinking several mm.He then attempted a size 6 broach and it countersunk as well.Lastly, he reimplanted the stem for the third time and it seated within a few mm of complete seating.There is no additional information available at the time of this report.
|
|