MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP PULSESELECT; PERCUTANEOUS CARDIAC ABLATION CATHETER FOR TREATMENT OF ATRIAL FIBRILLATION WITH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP PULSESELECT; PERCUTANEOUS CARDIAC ABLATION CATHETER FOR TREATMENT OF ATRIAL FIBRILLATION WITH Back to Search Results
Model Number PSCC100
Device Problems Material Integrity Problem (2978); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Loss of Vision (2139); Embolism/Embolus (4438); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/19/2024
Event Type  Injury  
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported that during a pulsed field ablation (pfa) procedure, fluoroscopy image showed that the array of the pfa catheter was inverted.The guidewire was pulled into the catheter and the slider was moved forward in order to try to move the array back to its shape.Additionally, noise from electrodes 1-2 was shown un the electrophysiology (ep) recording system.The catheter was not repla ced. the case was completed with pulsed field ablation.After removal of the catheter from the body, some damage was detected on the plastic between electrodes 1-2.It appeared to be kinked.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was later reported that no medical or surgical intervention was carried out but the patient's hospitalization was extended.The patient has not yet recovered from the embolism and vision issue.
 
Event Description
Additional information was received indicating that after the procedure and before discharge, the patient complained about an eye problem.The patient list their vision partially.Medical tests were performed.It was concluded that the patient suffered an embolism that affected the retinal vein.The patient was discharged.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSESELECT
Type of Device
PERCUTANEOUS CARDIAC ABLATION CATHETER FOR TREATMENT OF ATRIAL FIBRILLATION WITH
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18972942
MDR Text Key338594473
Report Number3002648230-2024-00109
Device Sequence Number1
Product Code QZI
Combination Product (y/n)N
Reporter Country CodeSP
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPSCC100
Device Catalogue NumberPSCC100
Device Lot Number0012090493
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/19/2024
Initial Date FDA Received03/25/2024
Supplement Dates Manufacturer Received04/05/2024
04/05/2024
04/12/2024
Supplement Dates FDA Received04/11/2024
04/11/2024
04/22/2024
Date Device Manufactured02/05/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-