ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUICKFLEX LV LEAD; Drug eluting permanent left ventricular (lv) pacemaker electrode Back to Search Results
Model Number 1156T
Device Problems Failure to Capture (1081); Fracture (1260); High impedance (1291)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Voluntary fda medwatch form received: mw5151638.Note: the lead's serial number and device details are unknown.
 
Event Description
It was reported that the system with a cardiac resynchronization therapy defibrillator (crt-d) and this left ventricular (lv) lead triggered an alert for lv automatic threshold test.It appeared that the test had not been able to be run for various days, nonetheless, when the crt-d was analyzed, it was no longer in suspension mode.The impedance values had increased several months ago, the behavior had been unstable and failure to capture was observed on the lv lead.Before the high, out of range pacing impedances the presenting electrogram appeared different and the lead was capturing.A lead fracture was suggested, and the health care professional agreed.The crt-d and this lv lead remain in service.No symptoms or adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKFLEX LV LEAD
Type of Device
Drug eluting permanent left ventricular (lv) pacemaker electrode
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18976342
MDR Text Key338593203
Report Number2017865-2024-35852
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1156T
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/06/2024
Initial Date FDA Received03/25/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-