• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. VISERA CYSTO-NEPHRO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. VISERA CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-VA2
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was observed that during the device inspection, the cysto-nephro videoscope exhibited shaving of forceps stopper mouthpiece.There were no reports of patient involvement.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.New information added to the following fields: h3, h6.Since the device is more than 15 years old, the device history record was unable to be reviewed.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.The device was returned, and an evaluation was completed for it.During inspection, olympus confirmed the reported event " the instrument channel port was scraped." based on the results of the investigation, a definitive root cause could not be determined.The event can be prevented by following the instructions for use which state: "visera cysto-nephro videoscope olympus cyf type va2 important information ¿ please read before use warnings and cautions ¿ do not twist or bend the bending section with your hands.Equipment damage may result.¿ do not squeeze the bending section forcefully.The covering of the bending section may stretch or break leading to fluid invasion into the endoscope.¿ do not bend the insertion section of the endoscope.The insertion section may be damaged.Chapter 7 cleaning, disinfection, and sterilization procedures 7.5 manual cleaning brushing the channel gently withdraw the single-use channel cleaning brush (bw-201b) from the suction cylinder opening part.If the shaft is rubbed strongly against the opening of the suction cylinder, the shaft may be shaved, and shavings may enter inside tubes.If cleaning is performed according to this instruction manual, shavings will not remain in the tube." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18976627
MDR Text Key339332007
Report Number3002808148-2024-02853
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170339455
UDI-Public04953170339455
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberCYF-VA2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/18/2024
Initial Date FDA Received03/26/2024
Supplement Dates Manufacturer Received04/08/2024
Supplement Dates FDA Received04/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/24/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-