Model Number BI70000028230
Device Problem Radiation Output Failure (4027)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/22/2024
Event Type  malfunction  
Manufacturer Narrative
References the main component of the system.Other medical products in use during the event include: brand name ; product id bi31000129 (lot: -); a medtronic representative went to the site to test the equipment.Testing revealed that the central processing unit (cpu) f luoro was replaced.The system then passed testing.Codes b01, c07 and d02 are applicable to this analysis.  medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that the 2d exposure failed, but 3d was normal. there was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM 3RD EDITION
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18977173
MDR Text Key338555237
Report Number3004785967-2024-00199
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K092564
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberBI70000028230
Device Catalogue NumberBI70000028230
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/22/2024
Initial Date FDA Received03/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/28/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse