• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL BRK¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH; TROCAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL BRK¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH; TROCAR Back to Search Results
Model Number G407208
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Great Vessel Perforation (2152)
Event Date 03/15/2024
Event Type  Injury  
Event Description
During an atypical flutter for a congenital heart (tridux surgery), the aorta was perforated.The introducer was removed, patient remained stable, and the procedure was completed.During the transseptal puncture, the aorta was perforated.Fluoroscopy confirmed the guidewire and dilator tip were in the aorta 1cm above the aortic valves.Cardiothoracic surgeons were called to advise.Structural abnormalities of the heart associated with previous surgical procedures cause the reflux to occur between the aorta and the right atrium and not towards the pericardium.The medical team decided to remove the sheath and guide.The patient's blood pressure was stable so the procedure was completed.Protamine was administered and the patient was transferred to the recovery room.The patient has since been discharged.There were no reported performance issues with any abbott device.
 
Manufacturer Narrative
Additional information: d9, g3, h2, h3, h6 the results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH
Type of Device
TROCAR
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key18977725
MDR Text Key338580291
Report Number3008452825-2024-00158
Device Sequence Number1
Product Code DRC
UDI-Device Identifier05414734205139
UDI-Public05414734205139
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberG407208
Device Lot Number10199961
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/15/2024
Initial Date FDA Received03/26/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/19/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/09/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Life Threatening;
-
-