BOSTON SCIENTIFIC CORPORATION TRAPEZOID RX; LITHOTRIPTOR, BILIARY MECHANICAL Back to Search Results
Model Number M00510870
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2024
Event Type  malfunction  
Manufacturer Narrative
Block h6: imdrf device code a0406 captures the reportable event of side car pushback.
 
Event Description
It was reported to boston scientific corporation that a trapezoid rx was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6)2024.During procedure, the side car was pushed back after the basket was pulled through the bile duct several times.A photo of the trapezoid rx provided by the customer shows the side car rx was pushed back.The procedure was completed with an alternative device.There were no patient complications as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID RX
Type of Device
LITHOTRIPTOR, BILIARY MECHANICAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18977854
MDR Text Key338587406
Report Number3005099803-2024-00927
Device Sequence Number1
Product Code LQC
UDI-Device Identifier08714729296386
UDI-Public08714729296386
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K040447
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00510870
Device Catalogue Number1087
Device Lot Number0032625848
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/04/2024
Initial Date FDA Received03/26/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/12/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-