Model Number 97715
Device Problems Failure to Interrogate (1332); Overheating of Device (1437); Communication or Transmission Problem (2896); Insufficient Information (3190)
Patient Problem Burning Sensation (2146)
Event Date 03/10/2024
Event Type  malfunction  
Event Description
Information was received from a patient regarding an external device.The reason for call was patient reported they were having trouble charging their implant and the issue began sunday.Patient reported they would get an error message and they would get no device found.Patient also stated the paddle got warm and almost hot where the cord went into the paddle.An email was sent to the repair department to replace the recharger.
 
Manufacturer Narrative
Section d references the main component of the system.Other medical products in use during the event include: brand name intellis; product id 97755, serial: (b)(6), product type: 0213-recharger.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Manufacturer Narrative
Continuation of d10: product id (b)(4) serial# (b)(6) product type recharger h3:analysis of the (b)(4) recharger (rtm) (serial number (b)(6) revealed a no device found message.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
Patient weight obtained.
 
Manufacturer Narrative
Continuation of d10: product id 97755 lot# serial# (b)(6) implanted: explanted: product type recharger medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18978634
MDR Text Key338598522
Report Number3004209178-2024-07816
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169781702
UDI-Public00643169781702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Remedial Action Notification
Type of Report Initial,Followup,Followup
Report Date 05/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/28/2021
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/12/2024
Initial Date FDA Received03/26/2024
Supplement Dates Manufacturer Received04/17/2024
04/17/2024
Supplement Dates FDA Received05/09/2024
05/10/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/07/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1535-2021
Patient Sequence Number1
Treatment
"SEE H11...."; "SEE H11...."; SEE H11...
Patient Age76 YR
Patient SexFemale
Patient Weight77 KG
-
-