Model Number 459888
Device Problems Failure to Capture (1081); High impedance (1291); Stretched (1601)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/29/2024
Event Type  malfunction  
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported that during the implant procedure, the physician went to slit the coronary sinus catheter and inner guide and the inner guide catheter was kinked and became more kinked when slitting was attempted.Upon trying to remove the catheter, the left ventricular (lv) lead was "stretched".The lead was tested and impedances were consistently high and was failing to capture.The lead was removed and a new lead was successfully implanted.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the full lead was returned and analyzed.No anomalies were found.The distal low voltage electrode of the lead was covered in body t issue/fibrotic growth.The analyst noted there were no anomalies observed regarding the returned lead.All electrical testing was within specified parameters.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN PERFORMA S MRI SURESCAN
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) P
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18979313
MDR Text Key338610624
Report Number2649622-2024-08217
Device Sequence Number1
Product Code OJX
UDI-Device Identifier00643169601871
UDI-Public00643169601871
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P080006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number459888
Device Catalogue Number459888
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/29/2024
Initial Date FDA Received03/26/2024
Supplement Dates Manufacturer Received04/03/2024
Supplement Dates FDA Received04/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexMale
-
-