• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX AFX2; BIFURCATED STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX2; BIFURCATED STENT GRAFT Back to Search Results
Model Number BEA28-110/I16-30
Device Problem Material Puncture/Hole (1504)
Patient Problem Aneurysm (1708)
Event Date 03/05/2024
Event Type  Injury  
Event Description
The patient was initially implanted with an afx2 bifurcated stent graft and a suprarenal aortic extension to treat an abdominal aortic aneurysm (aaa).Approximately six (6) years post initial procedure, evaluation of a computed tomography (ct) scan showed what appears to be a type 3b endoleak and the aneurysm measured 6.8 cm in diameter.The physician elected to reline the original implanted devices with another afx2 bifurcated stent graft and a vela infrarenal aortic extension to successfully resolve this event.The patient was reported as good post secondary procedure.
 
Manufacturer Narrative
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2 h3 other text : device remains implanted.
 
Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows that the type 3b endoleak of the distal main body and additional endovascular procedure are confirmed.This is consistent with the reported adverse event/incident.The clinical evaluation also shows reasonable evidence to suggest aneurysm enlargement of 12.2 mm occurred that was not included in the event as reported.The aneurysm enlargement was discovered during review of the computed tomography (ct) scan dated 03/25/2024.Procedure related harms, device, user, procedure or anatomy relatedness of this complaint could not be determined with the medical records available for review.The final patient status was reported as discharged home on postoperative day one in good condition.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2 g3: awareness date has been updated.H6: investigation finding codes: remove code 3233.H6: investigation conclusion codes: remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Type of Device
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key18979930
MDR Text Key338618445
Report Number3011063223-2024-00045
Device Sequence Number1
Product Code MIH
UDI-Device Identifier00818009014580
UDI-Public(01)00818009014580(17)180705
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/05/2018
Device Model NumberBEA28-110/I16-30
Device Lot Number1719186009
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/07/2024
Initial Date FDA Received03/26/2024
Supplement Dates Manufacturer Received04/26/2024
Supplement Dates FDA Received05/23/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/24/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0007-2019
Patient Sequence Number1
Treatment
AFX VELA SUPRARENAL, LOT # 1755304004
Patient Outcome(s) Required Intervention;
Patient Age74 YR
Patient SexMale
-
-