DEPUY IRELAND VELYS SAW INTERFACE LEFT; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 451570105
Device Problem Device Slipped (1584)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/22/2024
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi: (b)(4).
 
Event Description
It was reported that during device inspection it was found that the four (x4) robotic assisted saw interface devices 2 left and 2 right had clamps that were very loose and caused stability issues during cases.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.All available information has been disclosed.This is report 2 of 4 for (b)(4).
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: a video of the event was provided as well as the actual device.The video showed a left sasi assembled with a saw handpiece.The sasi was not assembled with the robot, however, in the video there was clearly some undesired movement or play between the sasi clamp and sasi itself while assembled with the saw handpiece.Visual analysis of the returned sasi found no significant damage or visual defects.The overall appearance of the device showed signs of normal wear from multiple uses in a clinical setting.The overall complaint was confirmed as the observed condition of the device would contribute to a device issue.The connection issues between the sasi and saw handpiece is related to design and has been escalated to a capa.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELYS SAW INTERFACE LEFT
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
DEPUY IRELAND
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic dr.
warsaw IN 46581
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key18980028
MDR Text Key338619267
Report Number1818910-2024-06696
Device Sequence Number1
Product Code OLO
UDI-Device Identifier10603295519461
UDI-Public10603295519461
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K202769
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number451570105
Device Lot NumberJ44924
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/22/2024
Initial Date FDA Received03/26/2024
Supplement Dates Manufacturer Received04/30/2024
05/21/2024
Supplement Dates FDA Received05/01/2024
05/21/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/24/2021
Type of Device Usage Unknown
Patient Sequence Number1
-
-