It was reported that the patient implanted with this right ventricular (rv) lead was recently seen in clinic and it was noted that loss of capture was observed.The right ventricular automatic thresholds were detected as greater than the programmed amplitude.The automatic threshold test was unable to collect data due to intrinsic beats/noise; however, the presenting electrogram shows clean artefact free channels.Additionally, the rv trend shows a sudden decrease in lead impedances from 776 ohms to 257 ohms.It was also noted that tachy mode is currently programmed to 'monitor' only.It is intended that a lead revision procedure will be performed.No adverse patient effects were reported.This lead currently remains implanted and in service.Additional information: a couple of days after the initial event notification, the field representative provided further information indicating that surgical intervention was performed to explant and replace this lead.It was also noted that a lead dislodgment occurred, which also led to high capture thresholds in addition to the other reported issues.The new implanted lead was positioned in the interventricular septum with good electrical measurements observed.There were no complications associated with the lead replacement procedure.No additional adverse patient effects were reported.This lead is not expected for return as it is currently in the possession of the hospital.
|