Model Number 419688 |
Device Problems
High impedance (1291); High Capture Threshold (3266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/29/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
It was reported that during the implant procedure, the left ventricular (lv) lead exhibited a high pacing threshold in the target vein, along with high pacing impedance values.A lead parameter test was performed a second time during the procedure, however, the high threshold and impedance continued to be observed.The lv lead was replaced with a left bundle branch block lead.No patient complications have been reported as a result of this event.
|
|