Catalog Number 4824062 |
Device Problem
Insufficient Information (3190)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2024 |
Event Type
malfunction
|
Event Description
|
The customer reported that during an apheresis platelet procedure on a trima device air was seen in the line going back to the donor during the first return.There were two small air bubbles, the first one was approximately 1/8th of an inch long and the second was approximately 1/8 inch long about 18 inches from the first air bubble.The operator clamped the line before air could be returned to the donor.All luer connections were tight and there was no clotting in the channel or return reservoir.The blood diversion pouch was not inflated with air and the donor was not connect prior to ac prime, and there was no air was being drawn in through the ac line or filter.No medical intervention was reported, and the patient was stable and was released home.Full donor unit id #:(b)(6) terumo bct is awaiting return of the disposable set for evaluation.
|
|
Manufacturer Narrative
|
Investigation: the run data file (rdf) was analyzed for this event.The analysis of the run data file did not identify a conclusive root cause for the presence of air in the donor/return lines for this procedure.No unusual process variable was identified and the signals in the run data file indicate that the trima accel system operated as intended.Specifically, the pressure signals and the pumped volume values were all within the expected ranges during tubing set test, ac prime, and blood prime states of the procedure.A review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.The customer history report indicates there were ten reports of a similar issue.A disposable complaint history search was performed for this lot and found two reports for similar issues on this lot.Investigation is in process, a follow-up report will be provided.
|
|
Search Alerts/Recalls
|
|
|