Model Number 8700-0650
Device Problem Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/12/2024
Event Type  malfunction  
Manufacturer Narrative
During preventative maintenance of the thermogard xp ivtm system sn (b)(6) by zoll at the customer site, the coolant pump stopped functioning.The root cause of the reported complaint was a cable contact failure on the coolant (march) pump.Upon visual inspection, no physical damage was noted.Upon servicing and replacing the coolant (march) pump, the thermogard xp ivtm system passed the final functional, calibration, and electrical safety tests without issues.Historical complaints were reviewed for service information related to the reported complaint, and there was no previous history of complaints reported for the thermogard xp ivtm system with sn (b)(6).
 
Event Description
During preventative maintenance of the thermogard xp ivtm system (sn (b)(6) by zoll at the customer site, the coolant pump stopped functioning.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZOLL IVTM THERMOGARD XP
Type of Device
THERMAL REGULATING SYSTEM
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kimthoa sackrison
2000 ringwood ave.
san jose, CA 95131
4084192922
MDR Report Key18984452
MDR Text Key338667372
Report Number3010617000-2024-00285
Device Sequence Number1
Product Code NCX
UDI-Device Identifier00849111075022
UDI-Public00849111075022
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K072234
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8700-0650
Device Catalogue Number8700-0650
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2024
Initial Date Manufacturer Received 03/12/2024
Initial Date FDA Received03/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/28/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-