MEDLINE INDUSTRIES, LP; DBD-ANGIO DRAPE CARDIAC PACK Back to Search Results
Model Number DYNJ51615K
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/28/2021
Event Type  malfunction  
Manufacturer Narrative
It was reported that a control syringe component was cracked.Reportedly, the problem/issue was identified during set up for a procedure.To date, no information has been received to indicate that a user or a patient experienced a death, serious injury, medical intervention, follow-up care, or other adverse health impact associated with the reported problem/issue.A photo was provided for review and the photographed control syringe had an area that appeared melted and displaced from the main barrel of the component.The are that appeared melted was observed to have holes at what appeared to be at equal intervals around the control syringe.No physical sample was returned for evaluation.In an abundance of caution, and in response to an fda 483 issued for cfn (b)(4) on 22-jan-2024, this medwatch is being filed for the reported problem/issue.If additional relevant information becomes available a supplemental medwatch will be filed.
 
Event Description
It was reported that a control syringe component was cracked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
DBD-ANGIO DRAPE CARDIAC PACK
Manufacturer (Section D)
MEDLINE INDUSTRIES, LP
three lakes drive
northfield IL 60093 2753
Manufacturer Contact
nigel vilches
three lakes drive
northfield, IL 60093-2753
MDR Report Key18984527
MDR Text Key338666927
Report Number1423395-2024-00221
Device Sequence Number1
Product Code OJG
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDYNJ51615K
Device Lot Number20EKA851
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/28/2021
Initial Date FDA Received03/27/2024
Was Device Evaluated by Manufacturer? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-