It was reported that a control syringe component was cracked.Reportedly, the problem/issue was identified during set up for a procedure.To date, no information has been received to indicate that a user or a patient experienced a death, serious injury, medical intervention, follow-up care, or other adverse health impact associated with the reported problem/issue.A photo was provided for review and the photographed control syringe had an area that appeared melted and displaced from the main barrel of the component.The are that appeared melted was observed to have holes at what appeared to be at equal intervals around the control syringe.No physical sample was returned for evaluation.In an abundance of caution, and in response to an fda 483 issued for cfn (b)(4) on 22-jan-2024, this medwatch is being filed for the reported problem/issue.If additional relevant information becomes available a supplemental medwatch will be filed.
|