SHIRAKAWA OLYMPUS CO., LTD. ULTRASONIC PROBE Back to Search Results
Model Number UM-S20-17S
Device Problem Stretched (1601)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/26/2024
Event Type  malfunction  
Event Description
It was reported that the ultrasonic probe tip was stretched.The issue occurred during preparation for use.The procedure was completed using a similar device.There were no reports of patient harm.
 
Manufacturer Narrative
To date, device has not yet been returned.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The returned device was evaluated and customer's allegation was confirmed.In addition, the device evaluation found indentation at the tip of the insertion.Based on the results of the investigation, "elongation of the tip of the insertion part" occurred due to external force on the tip of the insertion part.A definitive root cause for indentation of tip could not be identified.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Olympus will continue to monitor the field performance of this device.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the ultrasonic probe tip was stretched.The issue occurred during preparation for use.The procedure was completed using a similar device.There were no reports of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Type of Device
ULTRASONIC PROBE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18984672
MDR Text Key339053087
Report Number3002808148-2024-31688
Device Sequence Number1
Product Code ITX
UDI-Device Identifier04953170368479
UDI-Public04953170368479
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K982323
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUM-S20-17S
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 02/27/2024
Initial Date FDA Received03/27/2024
Supplement Dates Manufacturer Received04/06/2024
Supplement Dates FDA Received05/06/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/18/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-