Catalog Number RLT281416J
Device Problem Patient-Device Incompatibility (2682)
Patient Problem Aneurysm (1708)
Event Date 03/07/2024
Event Type  Injury  
Manufacturer Narrative
H3: code "other" was selected as the medical device remains implanted.Return not possible.H6: code d12: according to the gore® excluder® aaa endoprosthesis instructions for use (ifu), adverse events that may occur and/or require intervention include, but are not limited to: endoleak.Please note that the same patient was captured in medwatch report number 2017233-2024-04753.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
The following was reported to gore: on (b)(6) 2022, the patient underwent an endovascular treatment of an abdominal aortic aneurysm using gore® excluder® aaa endoprostheses.On (b)(6) 2022, the patient underwent an endovascular treatment of a thoracic aortic aneurysm using two gore® tag® conformable thoracic stent graft with active control system (tgmr373710j and tgm343415j).Tgmr373710j was implanted distally.On (b)(6) 2024, the patient was raced to the hospital for an impending rupture.It was confirmed that the impending rupture of the descending thoracic aortic aneurysm due to a distal type i endoleak from the gore® tag® conformable thoracic stent graft with active control system.And a proximal type i endoleak from the gore® excluder® aaa endoprosthesis was also observed.On (b)(6) 2024, a reintervention was performed.A gore® viabahn® endoprosthesis with propaten bioactive surface was implanted in the left renal artery and the right renal artery.A gore® tag® conformable thoracic stent graft with active control system was implanted to connect the previous implanted gore® tag® conformable thoracic stent graft with active control system and gore® excluder® aaa endoprosthesis (trunk ipsilateral leg).The procedure was concluded and the patient tolerated the procedure.The physician stated that the diameter of the infrarenal artery enlarged about 30 mm from about 22 mm and the diameter of the artery near the superior mesenteric artery enlarged about 34 mm from about 25 mm within 2 years after the gore® excluder® aaa endoprostheses and the gore® tag® conformable thoracic stent graft with active control system were implanted and these arteries enlargement led the endoleaks.And the physician also reported that the patient vessel condition was not well due to advanced age and it might be one of cause of the neck enlargement.
 
Manufacturer Narrative
H6: replaced c21 with c19: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Type of Device
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
asami aizawa
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key18984687
MDR Text Key338659134
Report Number3007284313-2024-03135
Device Sequence Number1
Product Code MIH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P020004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberRLT281416J
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/08/2024
Initial Date FDA Received03/27/2024
Supplement Dates Manufacturer Received03/08/2024
Supplement Dates FDA Received04/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/28/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age91 YR