ROCHE DIAGNOSTICS COBAS 8000 E 602 MODULE; IMMUNOCHEMISTRY ANALYZER Back to Search Results
Catalog Number 05990378001
Device Problems Low Test Results (2458); Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/19/2024
Event Type  malfunction  
Event Description
There was an allegation of questionable elecsys shbg results for 1 patient lithium heparin plasma sample on a cobas 8000 - cobas e 602 module.The initial shbg result was <0.8 nmol/l with flag.The customer was prompted to repeat the sample as it was accompanied by a data flag.The repeat result was 16.62 nmol/l.The repeat result was deemed correct.
 
Manufacturer Narrative
The reagent lot number and expiration date were not provided.The customer stated verbally that their qc was acceptable.The field service engineer (fse) observed multiple data flags on several different assays.The fse performed an instrument check and found that the sample probe was sticky along the shaft.The fse cleaned the sample probe and adjusted the voltage.The service maintenance actions performed by the fse resolved the issue.No further issues were reported after the service visit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS 8000 E 602 MODULE
Type of Device
IMMUNOCHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
HITACHI HIGH TECH CORP.
882 ichige hitachinaka
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18985253
MDR Text Key338748474
Report Number1823260-2024-00921
Device Sequence Number1
Product Code JJE
UDI-Device Identifier04015630929764
UDI-Public04015630929764
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K100853
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05990378001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/19/2024
Initial Date FDA Received03/27/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-