• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C8537
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/26/2024
Event Type  malfunction  
Manufacturer Narrative
E1: initial reporter address: (b)(6).G1: device manufacturer address 1: 600 mts.Oeste de entrada, principal ave.Las americas, parque industrial.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the tubing at the middle port of a clearlink system continu-flo solution set was completely detached.This was discovered before patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information: d9, h3, h4, h6, h10.H10: the device was received for evaluation.A visual inspection was performed and a separation between the tubing and the second y- site clearlink was observed in the upstream part.The reported condition was verified.The cause of the condition was determined to be lack of solvent at the tubing.The solvent was not applied uninformedly in the circumference during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18986798
MDR Text Key338711562
Report Number1416980-2024-01366
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412048970
UDI-Public(01)00085412048970
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K203609
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C8537
Device Lot NumberR23I30086
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/01/2024
Initial Date FDA Received03/27/2024
Supplement Dates Manufacturer Received04/26/2024
Supplement Dates FDA Received04/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NA.
-
-