BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 2C8537 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/26/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
E1: initial reporter address: (b)(6).G1: device manufacturer address 1: 600 mts.Oeste de entrada, principal ave.Las americas, parque industrial.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the tubing at the middle port of a clearlink system continu-flo solution set was completely detached.This was discovered before patient use.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
Additional information: d9, h3, h4, h6, h10.H10: the device was received for evaluation.A visual inspection was performed and a separation between the tubing and the second y- site clearlink was observed in the upstream part.The reported condition was verified.The cause of the condition was determined to be lack of solvent at the tubing.The solvent was not applied uninformedly in the circumference during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|