Catalog Number CDS0706-XTW
Device Problems Unintended Movement (3026); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/05/2024
Event Type  malfunction  
Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that a mitraclip procedure was performed to treat a functional mitral regurgitation (mr) with grade of 4.During the first establish final arm angle (efaa) step, the clip was observed to continuously open.It was indicated that when performing efaa, an audible click was heard.The clip arms were closed, and standard troubleshooting was performed.However, the clip was observed to continuously open during the second efaa.The clip was removed, and a replacement was used to complete the procedure.One clip was implanted, reducing mr to grade 1.There was no clinically significant delay in the procedure and no adverse patient effects.
 
Manufacturer Narrative
All available information was investigated, and the reported unintended movement (clip open ¿ efaa) and noise were not confirmed via returned device analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot-specific product issue.Based on the information provided and the results of the returned device analysis (unable to confirm the reported issues), a cause for the reported unintended movement (clip opened while establishing final arm angle) and noise-audible noise could not be determined.There is no indication of a product issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18987053
MDR Text Key338717604
Report Number2135147-2024-01360
Device Sequence Number1
Product Code NKM
UDI-Device Identifier05415067037381
UDI-Public(01)05415067037381(17)240917(10)30918R1003
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCDS0706-XTW
Device Lot Number30918R1003
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/05/2024
Initial Date FDA Received03/27/2024
Supplement Dates Manufacturer Received05/01/2024
Supplement Dates FDA Received05/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER
Patient Age83 YR
Patient SexMale
Patient Weight59 KG
-
-