Model Number R1054655TP
Device Problems No Display/Image (1183); Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/16/2024
Event Type  malfunction  
Event Description
The manufacturer received information alleging trilogy 100 screen displayed nothing and remained black.The family member stated airflow was delivered when the unit was turned on and that the issue occurred several times over several days.A sales rep confirmed the reported issue but could not confirm any alarms recorded in the event log.The rep did confirm circuit disconnect at suctioning and low minute ventilation in the event log.There was no harm or injury reported.The ventilator was returned to the manufacturer for evaluation and the customer¿s complaint was not duplicated therefore was not confirmed.Since a temporal failure could have occurred in lcd, lcd was replaced preventively to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY100
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18987071
MDR Text Key338717865
Report Number2518422-2024-15372
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00606959025387
UDI-Public00606959025387
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K083526
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberR1054655TP
Device Catalogue NumberR1054655TP
Initial Date Manufacturer Received 02/16/2024
Initial Date FDA Received03/27/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/27/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-