Catalog Number 8065753057 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Capsular Bag Tear (2639); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 02/26/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A customer reported that an ophthalmic operating system was used for the surgery.The patient experienced posterior capsule rupture and anterior chamber collapse during the cataract surgery for which the treatment was altered to vitrectomy.The current condition at the time of report was not recovered.
|
|
Manufacturer Narrative
|
The company representative was able to confirm the issue (via event log review) but was unable to replicate the reported event.The system was tested and found to meet product specifications.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A non-conformance based review of the batch/lot/serial number was performed and did not reveal a contributing factor.A review for complaints reported against this serial number was performed.A potentially relevant complaint was found and reviewed as part of this investigation.Posterior capsule tear is an issue that is occasionally reported with cataract surgery.However, a review of the complaint trends shows that the frequency reported is within known levels for this event.A root cause cannot be determined conclusively.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
|
Search Alerts/Recalls
|