Model Number PXMK2041
Device Problems Material Separation (1562); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/24/2023
Event Type  malfunction  
Manufacturer Narrative
The product is not expected to be returned for analysis since it was discarded.An engineering investigation will be performed in order to consider any potential factors that may have contributed to this complaint and a supplemental report will be sent with the investigation results.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
As reported, during use in patient in a life-threatening emergency, this disposable pressure transducer (dpt) (2024-06126-01) unscrewed when connected to the arterial access.It was necessary to redo the entire assembly of the pressure line by manually continuing the purge on the arterial line in order to prevent it from clogging.To solve the issue, a new device was opened (2024-06126-02), however, it presented the same issue as two of the screw threads were insufficiently screwed.All screw threads were checked before new connection.Patient demographics asked.There was no allegation of patient injury.The devices were not available for evaluation.
 
Manufacturer Narrative
Updated: h6 (device code, investigation findings, investigation conclusions), h10 (related report numbers).The manufacturing records were reviewed for the lot involved and there is no indication of a related nonconformance.All process parameters were met and inspections passed successfully.No root cause could be identified since no product sample nor objective evidence was provided for investigation.In addition, reviews of the device history records revealed no indication that a manufacturing non-conformance contributed to the reported complaint.Current risk mitigation(s) include visual inspection by manufacturing personnel, sampling leak test and sampling flow test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUWAVE DISPOSABLE PRESSURE TRANSDUCER
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
Manufacturer (Section G)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
Manufacturer Contact
samantha eveleigh
1 edwards way
irvine, CA 92614
9492503939
MDR Report Key18990248
MDR Text Key338900240
Report Number2015691-2024-02389
Device Sequence Number1
Product Code DXO
UDI-Device Identifier07460691951464
UDI-Public(01)07460691951464(17)240607(11)220608(10)64440036
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K222216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPXMK2041
Device Lot Number64440036
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/11/2024
Initial Date FDA Received03/28/2024
Supplement Dates Manufacturer Received05/10/2024
Supplement Dates FDA Received05/17/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-