It was observed that during the device inspection, the gastrointestinal videoscope exhibited foreign material at the plastic distal end cover and the light guide lens.Additionally, the channel-tube had foreign material, which appeared to be an endoscope accessory from the cleaning, disinfection, and sterilization (cds) process.There were no reports of patient involvement.
|
This report is being supplemented to provide additional information based on the legal manufacturer's device evaluation and investigation.Based on the results of the device evaluation, the reported event was confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 4 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the reported event could not be identified.The suggested events are detectable and preventable by following the instructions for use which state: -ifu states the detection method in gif/cf/pcf-190 series operation manual chapter 3 preparation and inspection -ifu states the preventative measures in gif/cf/pcf-190 series reprocessing manual chapter 5 reprocessing the endoscope olympus will continue to monitor field performance for this device.
|