• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SENSEONICS INC. EVERSESNE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SENSEONICS INC. EVERSESNE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 101967-750
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Discomfort (2330)
Event Date 07/30/2021
Event Type  Injury  
Manufacturer Narrative
This report is being submitted retrospectively as part of internal review.The insertion of new sensor was off label since it was inserted in a close proximity to the older sensor.This in turn created a linking issue for the user with the new sensor.Overall, the most likely root cause of this incident is due to the procedural error from the inserting physician.The newly inserted or the current sensor came out unfortunately while removing the old sensor due to their close proximity.
 
Event Description
Senseonics was made aware of an incident where newly inserted sensor came out along with the previous sensor due to the close proximity, while removing.This would require an additional procedure to get inserted with another sensor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSESNE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown 20876 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown 20876 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown 20876-7005
MDR Report Key18991912
MDR Text Key338886550
Report Number3009862700-2024-00224
Device Sequence Number1
Product Code QCD
Combination Product (y/n)Y
Reporter Country CodeGM
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date02/14/2021
Device Model Number101967-750
Device Catalogue NumberFG-4400-10-302
Device Lot NumberWP07957
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/10/2021
Initial Date FDA Received03/28/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/14/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-