MEDTRONIC NAVIGATION, INC STEALTHSTATION FLEXENT; EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT
|
Back to Search Results |
|
Model Number 9736242 |
Device Problem
Display or Visual Feedback Problem (1184)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/26/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: 9736226, software version #: 2.1.0 h3, h6: no products have been returned to medtronic for analysis.Codes b17, c20, and d15 are applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
Medtronic received information regarding a navigation system being used during a functional endoscopic sinus surgery (fess) procedure.It was reported that the system went black for about 10-15 seconds.The screen then came back on.There was a reported delay to the procedure of less than one hour.There was no reported impact to the patient.
|
|
Manufacturer Narrative
|
Additional information added to b5 and d10.Continuation of d10: product id: 9736217, lot number: unknown; product id: 9736218, lot number: unknown; product id: 9736217r, lot number: unknown.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
Additional information was received.The new computer (aio) when received would prompt the black background and then say enter password.Restarting the system and reseating the ssd did not resolve the issue.
|
|
Manufacturer Narrative
|
Product 9736218, lot number: s6pwnz0w502132t was received by the manufacturer for analysis.Reported problem could not be duplicated.System booted normally without failure.Touch screen functioned normally without failure.B01, c19, and d14 are applicable.Product 9736217r, lot number: 007-2390375 was received by the manufacturer for analysis.Reported problem was confirmed.Computer booted up in the bios screen and would not boot up into a known good ssd.The failure mode was electrical.B01, c02, and d02 are applicable.Case data was reviewed to determine if a software failure had occurred.It was determined no software failure occurred as the failure mode was an electrical hardware failure.B01, c19, and d14 are applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Manufacturer Narrative
|
The system was serviced in the field and the ssd was replaced.Codes b01, c02, and d02 apply.Product 9736217r was returned to the manufacturer unused and is not relevant to this event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|