Model Number A219 |
Device Problems
Failure to Sense (1559); Low impedance (2285)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/11/2024 |
Event Type
Injury
|
Event Description
|
It was reported that at a recent follow-up appointment the impedance of the subcutaneous implantable cardioverter defibrillator (s-icd) system was less than 30 ohms and there was a lack of sensing in the primary vector.X-ray revealed that the electrode had pulled back, resulting in sense b being in close proximity to the pocket.The patient underwent complete system replacement.No additional adverse patient effects were reported.
|
|
Event Description
|
It was reported that at a recent follow-up appointment the impedance of the subcutaneous implantable cardioverter defibrillator (s-icd) system was less than 30 ohms and there was a lack of sensing in the primary vector.X-ray revealed that the electrode had pulled back, resulting in sense b being in close proximity to the pocket.The patient underwent complete system replacement.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that the details of the event suggest the possibility that the patient twiddled the device system.
|
|