• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 36MM +5 OFFSET; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 36MM +5 OFFSET; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 6260-9-236
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Ambulation Difficulties (2544)
Event Date 02/28/2022
Event Type  Injury  
Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.H3 other text : device not returned.
 
Event Description
It was reported by claimant's counsel that allegedly the patient underwent right total hip arthroplasty on (b)(6) 2010, and was implanted with an lfit v40 femoral head and accolade tmzf stem.He underwent revision surgery on (b)(6) 2022, due to alleged dissociation of the femoral head from the trunnion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V40 COCR LFIT HEAD 36MM +5 OFFSET
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
anna ryan
raheen business park
limerick NA
EI   NA
61498200
MDR Report Key18992696
MDR Text Key338885735
Report Number0002249697-2024-00510
Device Sequence Number1
Product Code JDI
UDI-Device Identifier07613327032314
UDI-Public07613327032314
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K173499
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/31/2014
Device Catalogue Number6260-9-236
Device Lot NumberMHRHH3
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/29/2024
Initial Date FDA Received03/28/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/12/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number2249697-08/29/2016-007R
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age69 YR
Patient SexMale
-
-