Related manufacturer reference number:2017865-2024-37130, related manufacturer reference number:2017865-2024-37131.It was reported that patient presented to emergency room after getting inappropriate shock.Upon evalution, it was found from x-ray that patient's right ventricular (rv), left ventricular and atrial lead was dislodged.It was also noted that the lv lead exhibited loss of capture and atrial lead exhibited p wave amplitude variation.The rv and lv lead was explanted and replaced.The atrial lead was repositioned on (b)(6) 2024.The patient was stable.
|