MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH TELESCOPE, 70°, 4 MM; RIGID ENDOSCOPE TELESCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH TELESCOPE, 70°, 4 MM; RIGID ENDOSCOPE TELESCOPE Back to Search Results
Model Number A22003A
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/18/2024
Event Type  malfunction  
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.E1/establishment name: (b)(6), (added due to character limitation in respective field).
 
Event Description
It was reported, the telescope was cracked at the scope camera connection.The issue occurred during inspection for use of a therapeutic turis procedure.There were no reports of patient harm although there was a 3 minute delay.The procedure was able to be completed using a similar device.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was returned to olympus for inspection, and the customer's complaint was confirmed.Based on the results of the investigation, it is likely that the event occurred due to an effect of an excessive mechanical force caused by an impact or fall.However, the definitive root cause was unable to be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 70°, 4 MM
Type of Device
RIGID ENDOSCOPE TELESCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18993800
MDR Text Key338921358
Report Number9610773-2024-00895
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761020862
UDI-Public04042761020862
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K897003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22003A
Device Lot Number662575
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/21/2024
Initial Date FDA Received03/28/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/21/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
HD CAMERA HEAD
-
-