Model Number 3100 A
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/06/2024
Event Type  malfunction  
Event Description
It was reported to vyaire medical that audible internal leak occurred on the 3100 a device.The customer reported there was no patient involvement associated with the reported event.
 
Manufacturer Narrative
Vyaire medical file identification: (b)(4).H10: a vyaire field service representative(fsr) evaluated the device onsite and removed the top cover to look for any leaking parts.Nothing found at that time.2k pm was performed on the unit to ensure that the unit was working properly.After a few corrections on the unit it is ready for patient usage.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Manufacturer Narrative
H11: after further review of this complaint.Vyaire medical found out that this complaint is a duplicate of (b)(4).An original mdr for this reported issue was submitted under mfr report # 2021710-2024-18802.Please disregard and void mfr report #2021710-2024-18941.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
sandra valencia
510 technology drive
irvine, CA 92618
2402760001
MDR Report Key18996192
MDR Text Key338895081
Report Number2021710-2024-18941
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003079
UDI-Public(01)10846446003079
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100 A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/06/2024
Initial Date FDA Received03/28/2024
Supplement Dates Manufacturer Received05/07/2024
Supplement Dates FDA Received05/20/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-